Autoradiography is the direct exposure of film by beta particles or gamma rays. Fluorography is the exposure of film by secondary light that was generated by the excitation of a fluor or a screen by a beta particle or a gamma ray.
3H cannot be detected by autoradiography unless you have 1 mCi of it in your band. Because this is almost unheard of, it is essential to impregnate a gel containing 3H with PPO or Enlightening or to spray a 3H-containing TLC plate with spray Enhance or some such product in order to detect tritium.
35S and 14C can be detected by autoradiography. Simply expose your sample with unflashed film at room temperature or use the PhosphorImager.
32P and 125I can also be detected by autoradiography. Just expose your sample with unflashed film at room temperature.
Note that the image provided by direct autoradiography is always sharper than is the image resulting from fluorography.
Exposure with both isotopes can be accelerated greatly through the use of a screen.
When using a screen put flashed film between the screen and the gel or TLC plate and expose at -70°C.
Note that the film is exposed in two ways this way: directly by beta particles from the gel and secondarily by light from the screen. The majority of the exposure is from the screen. It is this side of the film that should be the so-called flashed side--see below.
Alternatively, use the PhosphorImager.
Exposure at -70°C does not enhance autoradiography. It does not hurt it however
Note: X-ray film has emulsion on both sides. It is useful therefore to keep track of which side of the film was up when the film was flashed. The film is quite opaque and the emulsion on the side facing the PPO'ed gel or the screen is exposed much more efficiently by the emitted light than is the other emulsion.
It seems likely in addition that flashing fogs the top emulsion more than the bottom emulsion. It is this side of the film that is referred to colloquially as the flashed side of the film. Therefore, to get the most out of flashing, the flashed side of the film should be put facing the PPO'ed gel or facing the screen, depending on which isotope is being detected.