摘要 目的:比较国内外不同厂家色谱柱对人胰岛素注射液的分离效果。方法:以不同配比的硫酸盐缓冲液和50
%乙腈溶液为流动相,检测波长为214 nm。结果与结论:通过调整流动相的不同配比,国内外不同厂家色谱柱均可使人胰岛素与脱酰胺胰岛素及防腐剂得到良好分离。 Suggestion and Evaluation of Chromatographic Columnsfrom Different Domestic and Foreign Manufacturers for the Separation of Human Insulin Injection by HPLC Zhang Ying and Chen Xiaoxia Abstract Objective:We make a comparision of columns between domestic and
Vydae for separation of human insulin injection by HPLC.Method:The mobile
phase is sodium sulfate buffer(pH 2.3)-50 % acetonitrile (45∶55)
or (47∶53).Detection wavelength is 214 nm.Result
and Conclusion:It has been shown
that columns from different domestic and foreign manufacturers could
be successfully applied to separate human insulin and its antiseptic
by HPLC after appropriately adjusting the proportion of the same
mobile phase. 基因重组人胰岛素自80年代研制成功以来,正在世界范围内逐步取代动物提取的胰岛素,其含量测定方法也由小鼠血糖法逐步被HPLC法所代替[1,2]。目前美、英、日及丹麦等国均用美国Vydae公司的ODS柱做人胰岛素含量测定,但该柱价格是国内ODS柱的8~10倍。本工作对国内的ODS柱是否可用于人胰岛素分析进行了考察,现报道如下: |
图1 在Vydac柱上混悬型样品的色谱图 |
图2 在YWG-C18柱上混悬型样品的色谱图
图注同图1
3.3 在YWG-C18柱上分离 YWG-C18柱系无定形填料及自行填充的色谱柱,当用流动相a时第二个防腐剂的峰与胰岛素峰的分离度小于1.5;用流动相b时,在20 min即可得到良好分离(图2)并可分离出脱酰胺胰岛素,对称性及分离度均符合药典要求。澄清型样品的色谱图同混悬型样品色谱图。 |
3.4 在天津特纳公司Kromasil ODS柱上的分离 该ODS柱用流动相a时约26
min得到分离(图3),峰的对称因子及分离度符合要求。但经调整流动相配比未能分离出样品中脱酰胺胰岛素及加入供试液中的脱酰胺胰岛素对照品。澄清型样品色谱图与混悬型样品色谱图一致。 |
图3 在特纳公司Kromasil ODS柱上混悬型样品色谱图
图注同图1
图4 在Hypersil ODS柱上混悬型样品的色谱图
图注同图1
4 讨论 * 日本JICA专家 参考文献 |