Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness

来源:英国眼科杂志
摘要:itAcceptedforpublication15June2003ABSTRACTPurpose:Toassessreproducibilityofcentralcornealthickness(CCT)measurementbymeansofultrasonicpachymetry。Methods:FiftyonevolunteersunderwentthreesessionsofCCTmeasurements,eachconsistingofthreeCCTmeasurements,performedbye......

点击显示 收起

1 Department of Ophthalmology, University of Milan Bicocca, Department of Neurosciences and Biomedical Technologies, Monza, Italy
2 Department of Ophthalmology, University of Milan, Department of Medicine, Surgery and Dentistry, Monza, Italy

Correspondence to:
Stefano Miglior
MD, Dipartimento di Neuroscienze e Tecnologie Biomediche, Universita’ degli Studi di Milano Bicocca, Via Cadore 48, 20052 Monza (MI), Italy; stefano.miglior@unimib.it

Accepted for publication 15 June 2003

ABSTRACT

Purpose: To assess reproducibility of central corneal thickness (CCT) measurement by means of ultrasonic pachymetry.

Methods: Fifty one volunteers underwent three sessions of CCT measurements, each consisting of three CCT measurements, performed by each of three different observers. Intra- and interobserver reproducibility was calculated by means of intraclass correlation coefficient (ICC). The expected range of variability between two independent evaluations was calculated using scatter plots of each test-retest difference against their mean. The standard deviation of the mean differences in the test-retest scores was used to describe the differences in the score spread.

Results: The ICC ranges of the intra- and interobserver evaluations were 0.95–0.97 and 0.89–0.95 respectively; the expected variability was ±1% and ± 2% respectively (95% confidence interval).

Conclusions: The measurement of CCT by means of ultrasonic pachymetry is highly reproducible.

Abbreviations: CCT, central corneal thickness; ICC, intraclass correlation coefficient; IOP, intraocular pressure; OHT, ocular hypertension

The measurement of central corneal thickness (CCT) is an important step in ophthalmic evaluations preceding kerato-refractive surgery.1 It is also an increasingly important procedure in the evaluation of patients with ocular hypertension (OHT) or glaucoma.2,3–20

CCT can be clinically assessed by means of optical or ultrasonic pachymetry,2 and optical coherence tomography.21 Only a few studies have attempted to evaluate the variability of ultrasonic pachymetry: their results show a good degree of reproducibility, although most of them involved small sample sizes or were designed to compare ultrasonic pachymetry with other methods of measuring CCT.22–29

The aim of this study is to evaluate the variability of ultrasonic pachymeter in the clinical setting and to provide a quantitative estimate of expected CCT measurements repeated by the same or different operators.

MATERIALS AND METHODS

One eye was randomly chosen for each of 51 volunteers aged 49–82 years (31 healthy individuals, 16 patients with OHT, and four patients with primary open angle glaucoma).

Individuals with previous corneal surgery, previous or current corneal disease, and contact lens wear were excluded from the study.

Ultrasonic pachymetry was performed with an undilated pupil using an "Altair" Ultrasonic Pachymeter (Optikon 2000, Rome, Italy) whose probe tip is approximately 1 mm in diameter.

The pachymeter was calibrated at the beginning of each session. After the instillation of a topical anaesthetic (oxibuprocain 0.4%), the probe was placed perpendicularly on the central cornea. This was confirmed by an audible beep produced by the instrument.

Three well trained operators (EA, MG, GM) independently measured the CCT of each eye in a random sequence within 3–4 minutes of each other in order to rule out the influence of possible diurnal variations in CCT.30–33 In order to reduce the possibility of ocular surface drying,27 one drop of artificial tear (Dacriosol, Alcon, Fort Worth, TX, USA) was instilled 30 seconds before each measurement. Each measurement was recorded by an assistant. The observers were masked to all CCT measurement.

Intraobserver reproducibility was calculated for each of the three examiners on the basis of three consecutive measurements.

Interobserver reproducibility was based on the analysis of the three independent series of measurements made by the three examiners (nine examinations).

Reproducibility was evaluated by means of the intraclass correlation coefficient (ICC).34–36

The ICC was calculated for each test-retest evaluation (E1-E2, G2-G3, etc, for the intraobserver assessment; E1-M2, M3-G1, etc, for the interobserver assessment), after which the mean ICCs were calculated with their ranges.

The expected range of variability between two CCT evaluations was calculated using the paired score differences of each test minus the retest. A scatter plot was then constructed by plotting each test-retest difference against its mean. The mean value and standard deviation (SD) of the test-retest score differences were then calculated, and the SD used to describe the spread of score differences. The 95% confidence intervals of the mean difference are thus the boundaries of the expected range of variability.37 Because the mean difference between the two repeated measurements of our 51 cases tended to be greater than zero, its value has been added to the standard calculation of 95% confidence intervals (which are also reported as per cent values around the mean measurement).

Statistical comparisons among groups were performed using Friedman’s non-parametric test; the statistical comparisons between two groups were made using Mann-Whitney’s non-parametric test.

Linear regression analysis was performed to evaluate whether there was a linear association between test-retest score difference and CCT value.

A variation of 15 μm between two repeated measurements has been considered as a clinically relevant outcome (as it can induce an error in intraocular pressure (IOP) assessment of about 1 mmHg according to the conversion factor of Ehlers and colleagues of 0.7 mmHg/10 μm4), and the frequency with which these substantial changes occurred has been calculated.

RESULTS

The descriptive statistics for each observer are summarised in table 1. The mean CCT measurements of the three observers were statistically different (p = 0.002).

   Table 1 Descriptive statistic of the CCT* measurements (μm) performed by the three observers

Intraclass correlation coefficient

The mean ICC for intraobserver reproducibility was 0.966 (SD 0.009) and ranged from 0.949 to 0.981. The mean ICC for interobserver reproducibility was 0.935 (SD 0.016) and ranged from 0.890 to 0.957. The ICCs of the intra- and interobserver analyses were significantly different (p <0.0001). The ICCs obtained by comparing observer M v E were not significantly different from those obtained by comparing observers M or E v G (p = 0.4).

Scatter plots

The expected intra- and interobserver measurements are shown in table 2. The mean expected intraobserver measurement was within ±0.75% and the largest observed value was < ±1.10%; the mean expected interobserver measurement was within ±1.20% and the largest observed value was < ±2%.

   Table 2 Scatter plots and mean expected range of variability from each test-retest evaluation

The mean expected measurements in the intraobserver and interobserver analyses were significantly different (p = 0.002). The mean expected measurements obtained by challenging observer M v E was significantly smaller than that obtained by challenging observers M or E v G (p<0.0001). Test-retest score difference did not regress on CCT value (r resulted less than 0.2 and p>0.1 in all analyses).

Clinically relevant changes

A variation of 15 μm between two repeated measurements occurred in 52 out of 459 (11.3%) test-retest intraobserver evaluations, and in 304 out of 1377 (22%) test-retest inter-observer evaluations.

DISCUSSION

There is considerable amount of published data suggesting a relationship between IOP and the risk of glaucoma,19,38–41 and CCT and the risk of glaucoma.2,20

We investigated the intra- and interobserver reproducibility of CCT measurements made using an ultrasonic pachymeter.

The descriptive statistics show that, although similar, the mean measurements made by the three observers were statistically different. However, this bias was extremely small as it could be estimated in only about 0.7% of the mean CCTs observed in our sample.

The analyses of intra- and interobserver reproducibility showed almost perfect agreement (ICC being between 0.81 and 0.99), and the mean expected variability found in the series of each single test-retest comparison was 1.92%.

Our results seem to indicate that both the intra- and inter-observer reproducibility of CCT measurements is extremely high. The fact that the instrument used in this study does not have a fixation light, and the fact that the probe diameter is only 1 mm (which may have induced some variation in its positioning on the cornea) may have contributed to variability in CCT measurements.

A further potential source of variability lies in the corneal touch technique. Our results in fact show that the CCT measurements made by different observers may be slightly different, and that the difference observed between the intra- and interobserver analyses was statistically significant.

The extent of variability between each test-retest did not depend on the absolute CCT value in any comparison, both in the intra- and interobserver evaluations.

A variation of ±15 μm between two repeated measurements occurred in 11.3% test-retest intraobserver evaluations, and in 22% test-retest interobserver evaluations. This indicates that, despite the high reproducibility of the procedure, care should be taken in the interpretation of IOP measurement corrected on the basis of CCT measurement. In fact, it is possible to expect 10% of the CCT measurements to induce an incorrect IOP estimate of about ±1 mmHg also when the same operator performs the CCT examinations.

Our study confirms the results of previous studies22–29 and provides a quantitative estimate of the CCT measurements that can be expected from repeat examinations.

The variability of IOP measurements made using the Goldmann applanation tonometer may be 2 mmHg in about 20–30% of subjects.42–44 If we assume that variations in CCT affect accurate applanation tonometry readings and apply Ehlers’ CCT correction factor (the largest reported so far) to more accurately determine IOP,4 the variability in ultrasonic pachymetry CCT measurements is less than the variability inherent in Goldmann tonometry IOP evaluations. This means that routine ultrasonic pachymetry CCT measurements should not introduce additional sources of error in clinical practice.

The results of our study suggest that the relatively simple, highly reproducible and objective nature of ultrasonic pachymetry should allow any well trained operator to make highly reliable CCT measurements.

REFERENCES

Carr J, Hersh P. Patient evaluation for refractive surgery. In: Azar DT, ed. Refractive Surgery. Stanford, CT: Appleton & Lange Publishers, 1997:101–9.

Doughty MJ, Zaman ML. Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol 2000;44:367–408.

Ehlers N, Hansen FK. Central corneal thickness in low-tension glaucoma. Acta Ophthalmol (Copenh) 1974;52:740–6.

Ehlers N, Bramsen T, Sperling S. Applanation tonometry and central corneal thickness. Acta Ophthalmol (Copenh) 1975;53:34–43.

Herndon LW, Choudhri SA, Cox T, et al. Central corneal thickness in normal, glaucomatous and ocular hypertensive eyes. Arch Ophthalmol 1975;115:1137–41.

Johnson M, Kass MA, Moses RA, et al. Increased corneal thickness simulating elevated intraocular pressure. Arch Ophthalmol 1978;96:664–5.

Whitacre MM, Stein RA, Hassanein K. The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 1993;115:592–6.

Argus WA. Ocular hypertension and central corneal thickness. Ophthalmology 1995;102:1810–12.

Wolfs RCW, Klaver CCW, Vingerling JR, et al. Distribution of central corneal thickness and its association with intraocular pressure. The Rotterdam study. Am J Ophthalmol 1997;123:767–72.

Stodtmaister R. Applanation tonometry and correction according to corneal thickness. Acta Ophthalmol Scand 1998;76:319–24.

Copt RP, Thomas R, Mermoud A. Corneal thickness in ocular hypertension, primary open-angle glaucoma and normal tension glaucoma. Arch Ophthalmol 1999;117:14–16.

Bron AM, Creuzot-Garcher C, Goudeau-Boutillon S, et al. Falsely elevated intraocular pressure due to increased central corneal thickness. Graefes Arch Clin Exp Ophthalmol 1999;237:220–4.

Shah S, Chatterjee A, Mathai M, et al. Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic. Ophthalmology 1999;106:2154–60.

Herman DC, Hodge DO, Bourne WM. Increased corneal thickness in patients with ocular hypertension. Arch Ophthalmol 2001;119:334–6.

Brandt JD, Beiser JA, Kass MA, et al. Central corneal thickness in the Ocular Hypertension Treatment Study (OHTS). Ophthalmology 2001;108:1779–88.

Goldmann H, Schmidt T. Uber applanationstonometrie. Ophthalmologica 1957;134:221–42.

Gloster J, Perkins ES. The validity of the Imbert-Flick law as applied to applanation tonometry. Exp Eye Res 1963;2:274–83.

Feltgen N, Leifert D, Funk J. Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings. Br J Ophthalmol 2001;85:85–7.

Gordon MO, Beiser JA, Brandt JD, et al. The Ocular Hypertension Treatment Study. Baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol 2002;120:714–20.

Medeiros F, Sample PA, Weinreb RN. Corneal thickness measuremens and visual function abnormalities in ocular hypertensive patients. Am J Ophthalmol 2003;135:131–7.

Muscat S, McKay N, Parks S, et al. Repeatibility and reproducibility of corneal thickness measurements by optical coherence tomography. Invest Ophthalmol Vis Sci 2002;43:1791–5.

Salz JJ, Azen SP, Berstein J, et al. Evaluation and comparison of sources of variability in the measurement of corneal thickness with ultrasonic pachymeters. Ophthal Surg 1983;14:750–4.

Villasenor RA, Santos VR, Cox KC, et al. Comparison of ultrasonic corneal thickness measurements before and during surgery in the Prospective Evaluation of Radial Keratotomy (PERK) study. Ophthalmology 1986;93:327–30.

Gordon A, Boggess EA, Molinari JF. Variability of ultrasonic pachymetry. Optom Vis Sci 1990;67:162–5.

Giasson C, Forthomme D. Comparison of central corneal thickness measurements between optical and ultrasound pachometers. Optom Vis Sci 1992;69:236–41.

Wheeler NC, Morantes CM, Kristensen RM, et al. Reliability coefficients of three corneal pachymeters. Am J Ophthalmol 1992;113:645–51.

Stucchi CA, Gennari G, Aimino G, et al. Systematic error in computerized pachymetry. Ophthalmologica 1993;207:208–14.

Yaylali V, Kaufmann SC, Thompson HW. Corneal thickness measurement with the Orbscan Topography System and ultrasonic pachymetry. J Cataract Refract Surg 1997;23:1345–50.

Marsich MW, Bullimore MA. The repeatability of corneal thickness measures. Cornea 2000;19:792–5.

Harper CL, Boulton ME, Bennet D, et al. Diurnal variations in human corneal thickness. Br J Ophthalmol 1996;80:1068–72.

Lattimore MR Jr, Kaupp S, Schallhorn S, et al. Orbscan pachymetry: implications of a repeated measures and diurnal variation analysis. Ophthalmology 1999;106:977–81.

Shah S, Spedding C, Bhojwani R, et al. Assessment of the diurnal variation in central corneal thickness and intraocular pressure for patients with suspected glaucoma. Ophthalmology 2000;107:1191–3.

Feng Y, Varikooty J, Simpson TL. Diurnal variation of corneal and corneal epithelial thickness measured using optical coherence tomography. Cornea 2001;20:480–3.

Fleiss JL, Cohen J. The equivalence of weighted kappa and the intraclass correlation coefficient as measures of reliability. Educational and Psychological Measurement 1973;33:613–19.

Snedecor GW, Cochran WG. Statistical Methods, 8th edn. Ames: Iowa State University Press, 1989:242–3.

Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics 1977;33:159–74.

Altman DG, Bland JM. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;8:307–10.

Leske MC, Connell AMS, Wu SY, et al. Incidence of open-angle glaucoma. The Barbados Eye Studies. Arch Ophthalmol 2001;119:89–95.

The AGIS Investigators. The Advanced Glaucoma Intervention Study (AGIS): 7. The relationship between control of intraocular pressure and visual field deterioration. Am J Ophthalmol 2000;130:429–40.

Kass MA, Heuer DK, Higginbotham EJ, et al. The Ocular Hypertension Treatment Study. A randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Arch Ophthalmol 2002;120:701–13.

Hejil A, Leske MC, Bengtsson B, et al. Reduction of intraocular pressure and glaucoma progression. Results from the Early Manifest Glaucoma Trial. Arch Ophthalmol 2002;120:1268–79.

Moses RA, Liu CH. Repeated applanation tonometry. Am J Ophthalmol 1968;66:89–91.

Phelps CD, Phelps GK. Measurement of intraocular pressure: a study of its reproducibility. Graefes Arch Clin Exp Ophthalmol 1976;198:39–43.

Sudesh S, Moseley MJ, Thompson JR. Accuracy of Goldmann tonometry in clinical practice. Acta Ophthalmol 1993;71:185–8.


 